VUV-14D Fixed Wavelength UV-VIS Absorbance Monitor
The VUV-14D is a fixed wavelength light absorbance monitor designed to be used with ultra low volume sampling cells under flowing solvent conditions (i.e. chromatography). The VUV-14D series of detectors has a high resolution analog output with a sensitivity range from 4.0 AU/V to 0.0005 AU/V and feature a state-of-the-art noise level of less than +/-0.00001 AU.
* Constant temperature with standard 10 mm cell (air).
Light source | Deuterium 190 nm - 400 nm |
Optical design | Double beam (sample/reference) |
Wavelength selection | Customer replaceable interference filter |
Wavelength accuracy | Defined by interferance filter (varies by wavelength) |
Wavelength precision | Better than +/- 0.2 nm |
Band pass | Deuterium lamp typically 10.0 nm at half height (dependent on interference filter) |
Noise* | <+/- 1 x 10-5 AU over 30 sec period peak to peak with 1.0 sec rise time |
Drift* | <+/- 1 x 10-4 AU over 30 min period peak to peak 1.0 sec rise time |
Input/Output | RS-232 computer I/O; auto-injection (event or run); auto-zero input; one analog integrator or chart recorder output, (voltage range default 1.0 V/AU); selectable analog range outputs of 4.0, 2.0, 1.0, 0.5, 0.2, 0.1, 0.05, 0.01, 0.005, 0.002, 0.001, 0.0005 AU/V with voltage range of +/- 4V at any scale |
Display | Absorbance (5 digit LED); wavelength (3 digit LED); analog output range (momentarily from absorbance display by pressing range switch); reference and sample light levels |
Size | 6.6" x 9.4" x 15" (HxWxD), Approx. 12 lb (shipping weight) |